Setting normalization and other options

This section teaches you how to set up normalization by a beamstop counter, and other options. It assumes you have completed Parts 1 (or 2) and 3.

  1. Open the Options window by selecting “Advanced Options” in the Options menu.

  2. In the window that shows up select the Image/Header Format section on the left. In the area on the right click the Load Image button.

    config_norm_load_image_png

  3. In the window that pops up, select the AgBeh_A1_43_001_0000.tiff file. Click the Open button.

    • Note: You can select any image of the appropriate type, not just the behenate.
  4. In the Image/Header Format window you should now see header values loaded into the list. Click the Apply button at the bottom of the screen.

    config_norm_apply_png

  5. Click on the Normalization section in the options list on the left.

  6. In the fields at the bottom of the Normalization panel, make sure “/” is selected in the left dropdown menu, and enter I3/200000 in the large field.

    • Note: It is typical in SAXS to normalize by the transmitted intensity. At the CHESS G1 beamline, the beamstop counter is name I3, which is why we are using that name in the normalization expression.

    config_norm_values_png

  7. Click the Calc button to evaluate the expression for the counter values loaded in the Image/Header Format tab. You should get a value of 0.02404.

  8. Click the Add button to add the expression to the normalization list.

  9. Make sure the “Enable Normalization” checkbox at the top of the page is checked.

  10. Click OK to exit the options window.

  11. In the file list, select the AgBeh_A1_43_001_0000.tiff file and click the Plot button. You will see a curve get plotted in the top panel of the Main Plot.

  12. Click on the manipulation tab. You will see a data item loaded in the manipulation list.

    config_norm_start_point1_png

  13. Adjust the start point for q Min to remove the points with zero value at the start of the curve (these are q points entirely in the mask). Set q Min so that the first point is the peak of the curve on the main plot. This should be around point 13 (depending on your mask).

    config_norm_start_point2_png

  14. Open the Options window as in Step 1.

  15. Click on the Calibration section in the options list on the left. Set “Start plots at q-point number” to the number you just found in Step 13.

    • Note: This makes it so that every curve loaded from now on will by default not display the first n points, which are covered by the beamstop.

    config_norm_start_point3_png

  16. RAW also allows you to add arbitrary metadata to your radially integrated files. One use case is to provide metadata keys for data deposition (e.g. in the SASBDB). To start, click on the Metadata section in the options list on the left of the Options window.

  17. In the ‘Key’ field enter ‘Instrument’. In the ‘Value’ field enter ‘MacCHESS BioSAXS beamline’.

    config_norm_metadata1_png

  18. Click the Add button.

    config_norm_metadata2_png

  19. Click the OK button to exit the options window and save your changes.

  20. You have configured everything necessary, and are now ready to save your settings. Go to the File menu and select “Save Settings”.

  21. Save the settings as SAXS.cfg.

  22. These settings can now be used to process images, and can be reloaded when you open RAW by selecting “Load Settings” from the File menu.